Convenient and efficient rotary evaporation
The R-300 meets the highest expectations in convenience and versatility. Its modular design allows extension of the Rotavapor® R-300 to a fully integrated system where a central interface regulates each component.
Where is the rotation speed displayed?
The rotation speed, as well as heating bath temperature and lift position, are shown on the integrated bath display. When using an interface those parameters are shown on both the bath and the interface.
Why should I buy plastic coated glassware?
The transparent plastic coating (P+G) improves the durability and stability of the glass. It also protects the operator from broken bits of glass as well as chemicals in case of glass breakage.
EnquiryRecommendation: Set the bath temperature to 50 °C. Applying the Δ 20° rule results in 50/30/10 °C.
Recommendation: Use standard position (25 °) for best efficiency without jeopardizing the sample.
Recommendation: Use 1.8 mm thick flasks (1 L) for best temperature exchange at high safety.
Recommendation: Use 250 to 280 rpm for maximum turbulence at high durability.
Recommendation: Select a flask that accommodates approximately twice the starting sample volume.